ISE Sodium Electrode Bio
Test kit for the Cedex Bio Analyzer
For use in quality control / manufacturing process only.

collapse
| -
ISE Sodium Electrode Bio material number and pack size: Material Number Pack Size 06437184001 1 piece -
Sensor for determination of Sodium.
The ISE Sodium Electrode Bio is intended for quantitative determination of sodium in aqueous solutions using the Cedex Bio Instrument.
The ISE unit uses flow-through ion-selective electrodes and a reference electrode with an open liquid junction. Each electrode has a membrane that is sensitive to a particular type of ion, in this case sodium ions. - The ISE Sodium Electrode Bio is intended for quantitative determination of sodium in aqueous solutions using the Cedex Bio HT Instrument.
The ISE unit in the instrument uses ion-selective electrodes - in this case sensitive to sodium ions - and compares the electical potential upon contact to the sample solution to the signal of a reference electrode. The signal difference is used to calculate the sodium concentration refering to the signals of calibrator solutions. - Sensor is function tested.